CNYST DH1200

CNYST DH1200 Digital Solid Density Measuring Densitometer User Manual

Eredua: DH1200

Produktua amaituview

The CNYST DH1200 Digital Solid Density Measuring Densitometer is a precision instrument designed for accurate density measurement of solid materials. It features an RS-232 interface and can handle samples with a maximum weight of 1200g and a weighing precision of 0.01g. This manual provides essential information for the setup, operation, and maintenance of your densitometer.

CNYST DH1200 Densitometer front view

1. irudia: Aurrealdea view of the CNYST DH1200 Digital Solid Density Measuring Densitometer.

Paketearen edukia

Before proceeding with setup, please verify that all components are present:

  • DH1200 Densitometer Main Unit
  • Measurement Cage Assembly (including upper and lower frames, transparent container)
  • Elikatze-moldagailua
  • Pintzak
  • Calibration Weight (e.g., 100g)
  • Termometroa
  • Small Parts (screws, mesh basket)
  • Erabiltzailearen eskuliburua (dokumentu hau)

1. bideoa: Kutxa ireki eta gehiagoview of the CNYST DH1200 Densitometer components.

Konfigurazio Argibideak

  1. Lekua: Place the densitometer main unit on a stable, level surface, free from vibrations and direct sunlight.
  2. Assemble Measurement Cage:
    • Carefully remove the transparent container and the black measurement frame from the main unit.
    • Ensure the transparent container is clean and dry before use.
    • Place the transparent container onto the designated area on the main unit's weighing platform.
    • Position the black measurement frame over the transparent container, ensuring it is securely seated.
  3. Potentzia-konexioa: Connect the provided power adapter to the densitometer's power input port and then plug it into a suitable power outlet.
  4. Hasierako piztea: Sakatu ON/OFF button to power on the device. Allow a few moments for the display to stabilize.
  5. Zero Kalibrazioa: With the measurement cage empty and assembled, press the ZERO button to tare the scale. The display should show "0.000g".
CNYST DH1200 Densitometer with measurement cage assembled

Figure 2: The densitometer with the measurement cage properly assembled.

Funtzionamendu-argibideak

The DH1200 densitometer can measure the density of solid materials using the Archimedes principle. Follow these steps for accurate measurements:

  1. Prestatu Sample: Ensure your solid sample is clean and dry. For porous materials, consider pre-treatment as per relevant standards.
  2. Measure Weight in Air:
    • Jarri sample on the upper platform of the measurement cage (above the water).
    • Wait for the reading to stabilize. Record this weight (W1).
  3. Likidoa prestatu: Fill the transparent container with distilled water or another suitable liquid (e.g., ethanol) to a level that fully submerges the sample without overflowing. Ensure no air bubbles are present.
  4. Measure Weight in Liquid:
    • Carefully transfer the sample to the lower platform of the measurement cage, ensuring it is fully submerged in the liquid and not touching the sides or bottom of the container.
    • Wait for the reading to stabilize. Record this weight (W2).
  5. Calculate Density: The densitometer will automatically calculate and display the density based on the measured weights and the density of the liquid used. Refer to the device's display for the calculated density value.
  6. Garbiketa: After measurement, remove the sample and liquid. Clean the measurement cage and transparent container thoroughly.
CNYST DH1200 Densitometer showing the weighing platform

3. irudia: hurbileko planoa view of the densitometer's weighing platform and control panel.

Mantentzea

  • Garbiketa: Regularly clean the densitometer's exterior with a soft, dry cloth. For the measurement cage and transparent container, use mild soap and water, then rinse thoroughly and dry completely. Avoid abrasive cleaners or solvents.
  • Biltegiratzea: Store the densitometer in a clean, dry environment, away from extreme temperatures, humidity, and direct sunlight. Keep the dust cover on when not in use.
  • Kalibrazioa: Periodic calibration by a qualified technician is recommended to ensure continued accuracy.

Arazoak konpontzea

ArazoaKausa posibleaIrtenbidea
Irakurketa okerrak
  • Azalera ezegonkorra
  • Air bubbles in liquid
  • Sample touching container
  • Kalibrazioa behar du
  • Jarri gainazal egonkor eta berdinean.
  • Remove all air bubbles from the liquid and sample.
  • Ziurtatu sample is freely suspended.
  • Perform calibration or contact support.
Pantailak "GEHIEGIZKO KARGA" erakusten duSample weight exceeds maximum capacity.Kendu sample immediately. Use a sample within the 1200g limit.
Gailua ez da pizten
  • Elektrizitate-egokigailua ez dago konektatuta
  • Korronte-hartunearen arazoa
  • Egiaztatu korronte-moldagailuaren konexioa.
  • Saiatu beste korronte bat.

Zehaztapenak

CNYST DH1200 Densitometer with technical specifications listed

Figure 4: Visual representation of key technical specifications.

EzaugarriBalioa
Neurketa-eremua0.001—99.999 g/cm³
Density Resolution0.001 g/cm³
Gehienezko pisua1200 g
Weighing Precision0.01 g
betetako estandarrakASTM D792, GB/T1033, GB/T2951, JIS K6268, ISO 2781, ISO 1183, GB/T208, GB/T5163, GB/T 1933
Elementuaren pisua9.9 kilo
Paketearen neurriak16.54 x 12.6 x 6.69 hazbete
Energia iturriaBattery Powered (Note: Unit also uses AC adapter as shown in video)
ZiurtagiriaISO
FabrikatzaileaCNYST

Bermea eta Laguntza

For warranty information and technical support, please contact the seller directly. The seller for this product is anhui east electronic. You can find their contact information on the product's Amazon page or through your purchase history.

Babes-planak eskuragarri daude estaldura luzatukoetarako:

  • 3 Urteko Babes Plana
  • 4 Urteko Babes Plana
  • Babes osoa (hileko plana)

For more details on these plans, please refer to the product listing on Amazon.com.

© 2023 CNYST. All rights reserved.