Apera Instruments AI531

Apera Instruments PH820 Benchtop pH Meter Kit

Argibide eskuliburua

Produktua amaituview

The Apera Instruments PH820 Benchtop pH Meter Kit is designed for reliable and accurate pH measurements in laboratory settings. It features advanced functionalities including intelligent calibration, data management, and a durable, water-resistant structure. This manual provides detailed instructions for proper use and care of your PH820 meter.

Apera Instruments PH820 Benchtop pH Meter Kit with electrode and stand

Figure 1: Apera Instruments PH820 Benchtop pH Meter Kit, showing the main unit, electrode, temperature probe, and electrode arm.

Key features of the PH820 include:

Konfigurazioa

Despaketatzea eta osagaien egiaztapena

Carefully unpack all components from the kit. Verify that the following items are present:

Konektatzeko osagaiak

Connect the electrode, temperature probe, and power adapter to the corresponding ports on the back of the PH820 meter unit. Ensure all connections are secure.

Back panel of Apera Instruments PH820 pH meter showing connection ports

2. Irudia: Atzealdea view of the PH820 meter, highlighting the REF, pH/mV, TEMP, RS232, and DC 9V input ports.

  1. Elektrodoen konexioa: Plug the LabSen 211 pH Electrode into the port labeled "pH/mV".
  2. Tenperatura-zundaren konexioa: Insert the temperature probe into the port labeled "TEMP".
  3. Potentzia-konexioa: Connect the power adapter to the "DC 9V" port and then to a power outlet.
  4. Electrode Arm Assembly: Attach the flexible electrode arm and holder to the side of the meter unit for convenient electrode placement during measurements.

Funtzionamendu-argibideak

Basic pH Measurement

  1. Power on the meter by pressing the power button.
  2. Garbitu elektrodoa ur destilatuarekin eta lehortu astiro-astiro.
  3. Immerse the electrode and temperature probe into the sampdisoluzioa. Ziurtatu elektrodoaren punta guztiz murgilduta dagoela.
  4. Wait for the reading on the large LCD to stabilize. The meter displays pH and temperature simultaneously.
  5. Record the stable reading.

pH Kalibrazioa

The PH820 supports 1 to 5 points automatic calibration. Regular calibration ensures measurement accuracy.

  1. Prepare the standard calibration solutions (4.00 pH, 7.00 pH, 10.01 pH are included).
  2. Sakatu CAL botoia sakatu kalibrazio moduan sartzeko.
  3. Rinse the electrode and immerse it into the first calibration solution (e.g., 7.00 pH).
  4. The meter will automatically recognize the solution and begin calibration. Wait for the stability icon to appear.
  5. Once stable, the meter will prompt for the next calibration point. Repeat the rinsing and immersion process for subsequent solutions.
  6. The self-diagnosis function will provide feedback on the calibration process and electrode condition, including slope data.
  7. Exit calibration mode once all desired points are completed.

Datuen kudeaketa

The PH820 allows for GLP data management, storing up to 500 groups of data. Data can be output via USB to a PC using the provided PC-Link software (Windows-based system only). Auto-timing data logging can also be configured.

ORP Measurement (Optional)

The PH820 can measure Oxidation-Reduction Potential (ORP) when an ORP electrode (e.g., 301Pt-C, sold separately) is connected to the pH/mV port. Refer to the ORP electrode's specific instructions for calibration and measurement procedures.

Mantentzea

Elektrodoen zaintza

Proper care of the LabSen 211 pH Electrode is crucial for maintaining accuracy and extending its lifespan.

Close-up of the LabSen 211 pH electrode

3. irudia: hurbileko planoa view of the LabSen 211 pH Electrode, showing its glass body and sensor tip.

Kontagailuen Zaintza

The PH820 meter unit is IP54 water-resistant, meaning it is protected from dust and splashing water. However, it should not be submerged. Keep the meter clean and dry. Use a soft, damp cloth to wipe the surface. Avoid using abrasive cleaners or solvents.

Arazoak konpontzea

The PH820 features a self-diagnosis function that can assist in identifying and fixing common problems, particularly during calibration. Refer to the meter's display for diagnostic messages.

ArazoaKausa posibleaIrtenbidea
Unstable ReadingsDirty electrode; air bubbles; insufficient sample volume; expired calibration.Clean electrode; ensure no bubbles; use adequate sample; recalibrate.
Kalibrazioaren hutsegiteaIncorrect buffer solution; contaminated electrode; electrode aging.Use fresh, correct buffers; clean electrode; check electrode slope data (self-diagnosis) for replacement indication.
Pantailarik/Energiarik ezPower adapter issue; loose connection.Check power adapter and outlet; ensure all cables are securely connected.
Irakurketa okerrakElectrode not calibrated; temperature not compensated; damaged electrode.Perform calibration; ensure temperature probe is functioning; inspect electrode for damage.

For persistent issues not resolved by the above, please contact Apera Instruments customer support.

Zehaztapenak

ParametroaBalioa
Modelo zenbakiaAI531
pHaren zehaztasuna±0.002 pH±1 digit
Kalibrazio-puntuak1 to 5 points (Auto.)
Datuak biltegiratzea500 groups (GLP compliant)
Datuen irteeraUSB (PC software support)
Uraren ErresistentziaIP54
Potentzia Sarrera120V AC 60Hz
Potentzia Irteera9V 300mA
Paketearen neurriak10.67 x 10.2 x 6.65 hazbete
Pisua6.69 kilo
ZiurtagiriakCE, RoHS, ISO 9001:2015

Bermea eta Laguntza

Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of production. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments website. You may also contact their customer service directly for assistance with product operation, troubleshooting, or parts replacement.