VWR 89202-918

VWR North America 89202-918 Circulating Open Bath Temperature Controller User Manual

Eredua: 89202-918

1. Sarrera

This manual provides essential instructions for the safe and efficient operation, installation, and maintenance of the VWR North America 89202-918 Circulating Open Bath Temperature Controller. Please read this manual thoroughly before using the device to ensure proper function and to prevent potential hazards.

2. Segurtasun informazioa

Beti beti segurtasun-neurri hauek:

3. Produktua amaitu daview

The VWR 89202-918 is a precision temperature controller designed for use with open baths, providing accurate and stable temperature regulation for various laboratory applications. It features an intuitive control panel and a robust heating and circulation system.

VWR 89202-918 Circulating Open Bath Temperature Controller main unit

1. irudia: Aurrealdea view of the VWR 89202-918 Circulating Open Bath Temperature Controller. This image displays the main control unit with its digital display and control buttons, along with the power cord extending from the base.

Heating element and circulator pump of VWR 89202-918

2. irudia: Azpialdea view of the VWR 89202-918 showing the heating coil and circulation components. This image highlights the internal heating element and the pump mechanism responsible for fluid circulation within the bath.

Product label for VWR 89202-918 showing model and electrical specifications

3. irudia: Product label detailing the VWR North American Cat. No. 89202-918, model MX20S135-V11B, serial number, and electrical specifications (60 Hz/A/V, 120V). This label provides critical identification and power requirements for the device.

3.1. Osagaiak

4. Konfigurazioa

  1. Despaketatzea: Carefully remove the controller from its packaging. Inspect for any visible damage.
  2. Muntaketa: Securely attach the controller to the side of your open bath using the integrated mounting clamp. Ensure the heating element and pump are fully submerged in the bath fluid, but the control head remains above the fluid level.
  3. Fluido maila: Fill the open bath with the appropriate heat transfer fluid to a level that covers the heating element and pump, but is below the minimum fill line indicated on the controller (if applicable) and well below the control head.
  4. Potentzia-konexioa: Connect the power cord to a grounded electrical outlet matching the voltage specifications on the product label (120V, 60Hz).

5. Funtzionamendu-argibideak

5.1. Piztu/Itzali

5.2. Tenperatura ezartzea

  1. With the unit powered on, the display typically shows the current bath temperature.
  2. Sakatu EZARTU button (or equivalent) to enter temperature setting mode. The setpoint temperature will flash.
  3. Erabili UP eta BEHERA arrow buttons to adjust the desired temperature setpoint.
  4. Sakatu EZARTU again to confirm the new setpoint. The unit will begin heating or cooling to reach the target temperature.

5.3. Circulation Control

The circulation pump typically operates continuously when the unit is powered on to ensure uniform temperature. Refer to the specific model's advanced settings for pump speed adjustments if available.

6. Mantentzea

7. Arazoak

ArazoaKausa posibleaIrtenbidea
Unitatea ez da pizten.No power supply; damaged power cord; internal fuse.Check power connection and outlet. Inspect power cord. Contact service if fuse replacement is suspected.
Temperature not stable or incorrect.Insufficient fluid level; contaminated fluid; sensor issue; external drafts.Ensure proper fluid level. Replace fluid if contaminated. Minimize external temperature fluctuations. Contact service if sensor is faulty.
Circulation pump not working.Blocked impeller; motor failure.Disconnect power and inspect impeller for obstructions. Contact service if motor failure is suspected.
Errore kodea bistaratu da.Sistemaren akats espezifikoa.Note the error code and refer to the full service manual or contact VWR technical support for specific guidance.

8. Zehaztapenak

9. Bermea eta Laguntza

For warranty information, technical support, or service inquiries, please contact VWR North America directly or refer to the official VWR website. Keep your purchase receipt and product serial number (found on the product label, see Figure 3) readily available when contacting support.

VWR North America Contact: us.vwr.com/store/

Erlazionatutako dokumentuak - 89202-918

Aurreview VWR Signature Circulating Baths with Advanced Programmable Temperature Controller Instruction Manual
This instruction manual provides comprehensive guidance for the installation, operation, and maintenance of VWR Signature Circulating Baths with Advanced Programmable Temperature Controllers, designed for precise laboratory temperature control.
Aurreview VWR MX Immersion Circulator Instruction Manual
Instruction manual for the VWR MX Immersion Circulator, detailing installation, operation, maintenance, troubleshooting, and technical specifications for precise temperature control in laboratory baths.
Aurreview VWR Signature Circulating Baths with Advanced Digital Temperature Controller Instruction Manual
This instruction manual provides detailed guidance on the installation, operation, safety, and maintenance of VWR Signature Circulating Baths equipped with an Advanced Digital Temperature Controller. Learn about model specifications, troubleshooting, and technical information for precise temperature control applications.
Aurreview VWR 12 and 18 Liter Linear Shaking Water Bath Instruction Manual
Comprehensive instruction manual for the VWR 12 and 18 liter linear shaking water baths. Covers safety precautions, setup, operation, specifications, maintenance, troubleshooting, warranty, and disposal.
Aurreview VWR General Purpose Water Baths Operator's Manual | Laboratory Temperature Control
Comprehensive operator's manual for VWR General Purpose Water Baths, covering installation, operation, maintenance, troubleshooting, and safety guidelines for laboratory incubation, inactivation, and agglutination procedures.
Aurreview VWR General Purpose Water Baths Operator's Manual
Operator's manual for VWR General Purpose Water Baths, detailing installation, operation, safety, maintenance, and troubleshooting for laboratory use.