Elitech IPT-100, IPT-100S Temperature and Humidity Data Logger

Segurtasun-argibideak
- To ensure proper installation and usage of this product, please read carefully and follow the guidelines below:
Bateria
- Use three AAA batteries. Do not use other types of batteries to avoid damaging the device or causing other malfunctions.
- Do not disassemble, crush, strike, or heat the batteries. Do not place them in fire, as this may cause the batteries to explode and trigger a fire.
Gailua
- Do not use the device in flammable or explosive gas environments to avoid the risk of explosion or fire.
- lf the device emits a burnt smell or other unusual odors, disconnect the power and contact the manufacturer or supplier immediately.
Neurriak
- If the device will not be used for an extended period, remove the batteries and store the device in a dry, cool environment.
- Users are not allowed to make any unauthorized changes to the device.
- Unauthorized modifications may affect the accuracy of the device or even cause damage.
- Do not use the device outdoors to avoid malfunctions caused by rain, lightning, or other adverse weather conditions.
- Use the device within its designated temperature and humidity range.
- Do not subject the product to heavy impact.
- Measurement values may be affected by the following factors:
- Tenperatura-errorea:
- Insufficient stabilization time in the measurement environment.
Proximity to heat or cold sources or direct exposure to sunlight.
- Insufficient stabilization time in the measurement environment.
- Hezetasun errorea:
- Insufficient stabilization time in the measurement environment.Long-term exposure to steam, mist, water curtains, or condensation environments.
- kutsadura:
Presence of dust or other pollutants in the environment.
Produktua amaitu daview
- The temperature and humidity meter is designed for industrial workshops, warehousespharmacies, laboratories, agricultural environments, and other professional settings for monitoring environmental temperature and humidity.
Produktuaren Ezaugarriak
- Supports OTA updates for continuous improvements.
- Datuak izan daitezke viewed and traced through the mobile app for easier tracking.
- Scenario-based configuration with smart alerts.
- 4-inch large screen for clear data display.
Produktuen Itxura
- Max / Min botoia
- ℃/℉ toggle button
- Mode/Bluetooth button
- External probe interface
- Imana
- Bateriaren estalkia
- Zutik
- External temperature sensor (Standard for IPT-100S)
Produktuaren interfazea
- Bateriaren mailaren ikonoa
- Bluetooth ikonoa
- Placeholder icon (no function)
- Alarma ikonoa
- Recording display icon
- External probe display mode
- Max/Min indicator icon
- Denbora bistaratzeko eremua
- Backtrack temperature display area
- Backtrack time indicator area
- Tenperatura bistaratzeko eremua
- Over-limit status display (over upper limit
, over lower limit
- Humidity range indicator area
- Hezetasuna erakusteko gunea
Parametro Teknikoak
|
Neurketa-eremua |
Tenperatura:-10°C~50°C |
| Humidity:10%RH~99%RH | |
|
Zehaztasuna |
Temperature:+0.5°C(10-35°C),+1°C for other ranges |
| Humidity: +5%RH(40%-75%),+10% for other ranges | |
|
Ebazpena |
Tenperatura: 0.1°C |
| Hezetasuna: 0.1% RH | |
| Elikatze Hornidura | 3*AAA |
| Egonean Denbora | Not less than 6 months |
| Pantaila | 4.5-inch segmented display |
| Datuen Grabaketa | Not less than 5000 entries |
| Bluetooth barrutia | Up to 30 meters in an open environment |
| Ingurune Operatiboa | -20~60℃, 0~90%RH (non-condensing) |
| Biltegiratze ingurunea | -25~65℃, 0~90%RH (non-condensing) |
| Neurriak | 118.6*99.5*23.0mm |
Accessories External Probe
- (Applicable only to IPT-100S with built-in external probe)
Neurketa-eremua Temperature:-20°C~70℃ Zehaztasuna Temperature:+0.5°C(0℃~40°℃),+1°C for other ranges Ebazpena Tenperatura: 0.1℃ Interfaze mota Type-C port (Note: This port does not support charging!) Ingurune Operatiboa -40~85℃, 0~90%RH (non-condensing) Biltegiratze ingurunea -50~90℃, 0~90%RH (non-condensing) Sentsore mota Tenperatura sentsore digitala Lerroaren luzera 2 metro
Produktuaren Ezaugarriak
- Produktuen aktibazioa
- Remove the insulating strip from the back of the product to power it on. The screen will display all data for 2 seconds and then show the real-time readings from the internal sensor.
- Note: After powering on, connect the device to your phone via Bluetooth to sync the time. If not in use for a long period, remove the batteries and store them separately.
- The Type-C interface is only for connecting the external probe, not for charging.
- .Button Functions:
Botoia
Prentsa laburra
Sakatu luzea 2s

Switch between max, min, and current values Clear maximum and minimum values 
Toggle backtrack time Switch temperature unit (°C/°F)
Switch between internal/external probe data displayGaitu / Desgaitu Bluetooth konexioa - Modu aldaketa:
Press the mode button to switch between the internal sensor’s temperature and humidity data and the external probe’s temperature data.- Note: If no external probe is connected, the external probe mode will show “—” and the external probe icon will light up.
- View Max/Min Values:
short Press the max/min button to cycle between the maximum, minimum, and current values in the current mode.- Long-press the button to reset the maximum and minimum values.
- Humidity Indicator:
- By default, when the humidity is below 40%, the indicator turns red to show dry conditions. Between 40%-60%, it turns green to show comfortable conditions, and above 60%, it turns blue to show humid conditions. The app allows customization of these ranges.
- Note: Default values are based on the “Indoor Air Quality Standard” (GB/T 18883-2022)
- Product Alarms
Bistaratzea
Arrazoia Irtenbidea Bip Current data exceeds the limit Automatic mute after 60 seconds, or press any key to mute 
Current data exceeds muga
Automatically cancels when data normaltasunera itzultzen da

Exceeded the upper limit of the display range Automatically cancels when data returns to normal 
Exceeded the lower limit of the display range Ez da sentsorea hauteman
Automatically cancels when data returns to normal Insert or replace the sensor

APParen funtzionamendurako jarraibideak
APP konexioa
- Scan the QR code below to download and install the dedicated APP, or search for “Elitech Tools” in the app store to install.

- Luze sakatu
Bluetooth button for 2 seconds to enable Bluetooth. The Bluetooth icon will flash. Open the “Elitech Tools” app, search for nearby devices, find the correct model, and click to connect.
-
- Note: Ensure your phone’s Bluetooth is enabled while using the app. The device’s Bluetooth will automatically turn off if not connected within 2 minutes.
APP interfazearen aurkezpena
- The main interface displays the current built-in temperature, humidity, probe temperature, and max/min value. By clicking
, you can reset the max/min values. When the temperature or humidity exceeds the upper or lower alarm limits, the alarm icon will appear in the corresponding alarm zone
- Click the text in the upper right corner to switch Scene Mode
- Click the dial to switch to the curve interface and view the temperature and humidity curves.
- Click “Settings” to modify relevant parameters

- Click “Scene Settings” to select the scene you want to modify
- Modify the alarm parameters for this scene.
- Modify humidity range value
Eranskinen zerrenda
- Recording-type Temperature and Humidity Meter *1
- AAA Batteries *3
- Imanak*2
- External Temperature Sensor Probe *1
- (only for IPT-100S with built-in external probe)
Ohiko galderak
G: Gailua kanpoan erabil daiteke?
A: No, the device should not be used outdoors to avoid malfunctions caused by adverse weather conditions.
G: Nola kalibratu dezaket gailua?
A: Calibration is not user-accessible. Please contact customer Support if you believe your device requires calibration.
Dokumentuak / Baliabideak
![]() |
Elitech IPT-100, IPT-100S Temperature and Humidity Data Logger [pdf] Argibideen eskuliburua IPT-100, IPT-100S, IPT-100 IPT-100S Temperature and Humidity Data Logger, IPT-100 IPT-100S, Temperature and Humidity Data Logger, Humidity Data Logger, Data Logger, Logger |