Apera Instruments AI311

Apera Instruments AI311 PH60 Waterproof pH Pocket Tester Kit User Manual

Brand: Apera Instruments

Modeloa: AI311

1. Sarrera

The Apera Instruments PH60 Pocket pH Tester is designed for reliable and easy pH measurement in general water solutions. This includes applications such as hydroponics, horticulture, aquaculture, pools & spas, water treatment, cooling towers, environmental monitoring, beverage making, and educational settings. The meter features a replaceable pH probe (PH60-E) specifically for general water solutions. It is also compatible with three other types of pH electrodes and an ORP probe (sold separately) for specialized applications like solid samples and surface tests.

Oharra: A few drops of 3M KCL liquid (included in the kit) should be added to low ion-strength test samples (e.g., distilled water or RO water) to significantly accelerate stabilization without affecting the pH reading.

Close-up of the PH60 pH sensor bulb

Irudia: Gertuko planoa view of the high-quality glass pH sensor bulb, designed for accuracy and stability.

Infographic showing various applications for the PH60 pH tester

Image: Visual representation of diverse applications including hydroponics, pools, brewing, drinking water, water treatment, aquariums, and aquaculture.

2. Zer dago kutxan

Your Apera Instruments PH60 Waterproof pH Pocket Tester Kit includes the following items:

Apera Instruments PH60 pH Tester Kit contents laid out in its carrying case

Image: The complete PH60 pH Tester Kit, showing the meter, calibration solutions, bottles, and accessories neatly organized in its hard-shell case.

3. Konfigurazioa

3.1 Bateriaren instalazioa

The PH60 tester requires 4 AAA batteries (included). To install or replace batteries:

  1. Unscrew the battery cap at the top of the meter.
  2. Sartu 4 AAA pilak, polaritate zuzena (+/-) ziurtatuz.
  3. Itxi berriro bateriaren tapa ondo torlojuekin.
PH60 pH tester with battery compartment open, showing AAA batteries

Image: The top section of the PH60 pH tester with the cap removed, revealing the battery compartment for AAA batteries.

3.2 Initial Probe Preparation

Before first use or if the tester has not been used for a long time (over 1 month):

  1. Pull out the battery insulation slip (if present) and remove the probe cap.
  2. Rinse the probe thoroughly with pure water (preferably distilled or deionized water, RO water is acceptable). Shake off excess water.
  3. If the tester hasn't been used for a long time (over 1 month), soak the probe in the 3M KCL soaking solution (included in the kit) for 15 minutes.
  4. Proceed to calibrate the meter before testing.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This video demonstrates the initial preparation steps, including removing the battery insulation slip, rinsing the probe, and soaking it in KCL solution if necessary (0:13 - 0:55).

4. Funtzionamendu-argibideak

4.1 pH Kalibrazioa

The PH60 tester features 1-3 point automatic calibration with auto-buffer recognition. Always calibrate pH 7.00 first. If you need to calibrate pH 4.00 or pH 10.01, do so immediately after pH 7.00 calibration. Do NOT turn the meter off between pH 7.00 and pH 4.00/10.01 calibrations.

  1. Pour pH buffer solutions (pH 4.00, pH 7.00, pH 10.01) into the corresponding calibration vials to about half volume.
  2. Garbitu zunda ur puruan eta astindu gehiegizko ura.
  3. Short press the power button to turn on the tester.
  4. Dip the probe into the pH 7.00 buffer solution. Make a quick stir, then let it stand.
  5. Long press the 'CAL' button to enter calibration mode. The screen will turn green.
  6. Wait for the reading to stabilize (a smiley icon will appear on the screen), then short press 'CAL' to finish the first point calibration. The tester will return to measurement mode, and an 'M' icon will appear at the bottom left, indicating a successful first-point calibration.
  7. Without turning off the tester, rinse the probe and repeat steps 4-6 for pH 4.00 buffer solution. An 'L' icon will appear for successful 2-point calibration.
  8. Without turning off the tester, rinse the probe and repeat steps 4-6 for pH 10.01 buffer solution. An 'H' icon will appear for successful 3-point calibration.
Step-by-step guide on how to calibrate the PH60 pH tester

Image: Detailed infographic illustrating the 3-point calibration process for the PH60 pH tester.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates the pH calibration process (0:56 - 4:26).

4.2 pH-aren neurketa

To take a pH measurement:

  1. Short press the power button to turn on the tester.
  2. Garbitu zunda ur puruan eta astindu gehiegizko ura.
  3. Insert the probe into your sample solution, make a quick stir, and hold still.
  4. Take the reading as the measurement when it is fully stabilized (a smiley icon appears and stays on screen).
  5. If the Auto-Hold function is turned on (in parameter setting P4), the reading will be automatically locked when it's stable for more than 10 seconds. The 'HOLD' icon will stay on. Short press the 'CAL' button to cancel Auto-Hold and continue measuring.
  6. Short press the 'MODE' button to switch to mV mode, which is for ORP testing or technical analysis (an ORP probe is needed for ORP measurements).
Hand holding the PH60 pH tester, showing a pH reading of 7.00

Image: A hand holding the PH60 pH tester, displaying a pH reading of 7.00 and a temperature of 25.0°C.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates the pH measurement process (4:26 - 5:33).

5. Mantentzea

5.1 Zundaren garbiketa

The accuracy of the tester depends on the cleanliness of the probe. Always thoroughly rinse off the probe before and after each measurement with pure water.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates probe cleaning procedures (5:33 - 6:36).

5.2 Zundaren biltegiratzea

Proper storage is crucial for maintaining probe longevity and accuracy.

IMPORTANT: NEVER store the probe in pure water (like tap, RO, distilled, or deionized water) as they could damage the pH sensor. If this happens, immediately soak the pH probe in the 3M KCL soaking solution overnight to restore. Then re-calibrate it before using. Pure water is only for rinsing off the probe, NOT for storage.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates proper probe storage (6:36 - 7:29).

6. Arazoak

Hona hemen ohiko arazo batzuk eta haien konponbideak:

7. Zehaztapenak

EzaugarriBalioa
Produktuaren neurriak7 x 1.6 x 1.6 hazbete
Elementuaren pisua10 ontza
Elementuaren eredu-zenbakiaAI311
Pilak4 AAA pila behar dira (barne)
pHaren zehaztasuna±0.01 pH
pH barrutia-2.00 eta 16.00 pH
Tenperatura tartea32 eta 122 °F (0 eta 50 °C)
Itzalketa automatikoa8 minutu funtzionamendurik gabe

8. Bermea eta Laguntza

Apera Instruments is committed to providing reliable products and ongoing support to ensure you achieve the most accurate results. For warranty information, technical assistance, or any product-related inquiries, please refer to the official Apera Instruments website or contact their customer support directly. Contact details are typically provided on the product packaging or within the full user manual.

Erlazionatutako dokumentuak - AI311

Aurreview Apera PH60 pH neurgailuaren erabiltzailearen eskuliburua - Gida eta zehaztapenak
Apera Instruments PH60 Premium pH neurgailuaren erabiltzailearen eskuliburu osoa, konfigurazioa, kalibrazioa, neurketa, arazoak konpontzea eta zehaztapenak biltzen dituena.
Aurreview Apera EC20 poltsikoko eroankortasun-probatzailea: argibideen eskuliburua eta gida
Apera EC20 poltsikoko eroankortasun-probatzailearen instrukzio-eskuliburu osoa. Ikasi konfigurazioari, kalibrazioari, eroankortasunaren neurketari, zehaztapen teknikoei eta bermeari buruzko informazioari buruz.
Aurreview Apera Instruments PH20 Balio pH Tester Kit-aren Erabiltzaile Eskuliburua
Apera Instruments PH20 Value pH Tester Kit-erako erabiltzailearen eskuliburu osoa, bateriaren instalazioa, teklatuaren funtzioak, kalibrazioa, neurketa, parametroen ezarpenak, arazoak konpontzea, zehaztapen teknikoak eta bermearen informazioa biltzen dituena.
Aurreview Apera Instruments 5052 Seawater Salinity Pocket Tester Operation Manual
Detailed operation manual for the Apera Instruments 5052 Seawater Salinity Pocket Tester, covering technical parameters, calibration, measurement procedures, important notes, and warranty information.
Aurreview Apera PH60-HT pH neurgailuaren erabiltzailearen eskuliburua: Tenperatura altuko eta soluzio kaustikoak
Comprehensive user manual for the Apera PH60-HT pH Meter, designed for accurate pH measurement in high temperature (up to 140°C) and caustic solutions. Covers kit contents, operation, calibration, electrode care, parameter settings, technical specifications, troubleshooting, and warranty.
Aurreview Apera Instruments PH60 Premium pH neurgailuaren erabiltzailearen eskuliburua
Apera Instruments PH60 Premium pH neurgailuaren erabiltzailearen eskuliburu osoa, konfigurazioa, kalibrazioa, neurketa, mantentze-lanak, arazoak konpontzea eta zehaztapen teknikoak biltzen dituena.